Uploaded on Apr 10, 2025
MedicalDeals' Anesthesia Machine delivers precise O2 and N2O control with a pressure range of 280-600 kPa. Featuring volume control, pressure limit, and versatile settings for both adult and pediatric patients. Equipped with alarms and monitoring to ensure enhanced patient safety.
Anesthesia-Machine MD-AM-1000
Anesthesia Machine MD-AM-1000
Overview :
Anesthesia Machine MD-AM-1000 operates on a gas source of O2 and N2O, with a pressure range of 280 kPa to 600 kPa. It has
multiple ventilation modes, including IPPV, A/C, SIMV, SIGH, and MANUAL. It offers versatile working modes such as volume
control and pressure limit, which adjust according to the patient's condition. The machine supports both adult and pediatric patients.
Our Anesthesia Machine features various alarms and monitoring capabilities, to ensure patient safety.
Features:
7-inch TFT LCD screen display
Anesthesia Circuit Gas Outlet (ACGO) feature
Oxygen flush technique
High-precision flowmeter
4-tube mechanical flowmeters
Multiple working modes
Alarms for power, voltage, volume, and pressure
Applications:
Our Anesthesia Machine is used in the ICU and operating room to deliver general anesthesia to patients undergoing surgeries or
medical procedures.
Specifications :
Ventilation Mode IPPV, A/C, SIMV, SIGH, MANUAL
Ventilator Flowmeter O? (0.1 to 10 L/min) N?O (0.1 to 10 L/min)
Ventilator Rapid Oxygen Supply 25 L/min to 75 L/min
Ventilator Tidal Volume 0.20 mL to 1500 mL
Ventilator Frequency 1/min to 100/min
Ventilator Pressure triggering
-20 cm H?O to 0 cm H?O (Based on PEEP)
Sensitivity
Ventilator PEEP 0 cm H?O to 30 cm H?O
Ventilator Flow Trigger Sensitivity 0.5 L/min to 30 L/min
Ventilator SIGH 0 (off) 1/100 to 5/100
Ventilator Apnea Ventilation OFF, 5sec to 60 sec
Ventilator Pressure Limit 20 cm H?O to 100 cm H?0
Monitoring Frequency 0 /min to 100 /min
Monitoring Volume 0 mL to 2000 mL
Monitoring MV 0 L/min to 100 L/min
Monitoring Oxygen Concentration 15% to 100%
AC power failure Alarm Power failure or no connection
Low Voltage alarm for backup battery 11.3 ± 0.3 V
Volume (Tidal) Alarm ? 5 mL within 6s
Oxygen Concentration Alarm 19% to 100% (HIGH) 18% to 99 % (LOW)
Airway Pressure Alarm 20 cm H?O to 100 cm H?O (HIGH) 0 cm H?O to 20 cm H?O (LOW)
Adult (5 L/min to 20 L/min) Child (1 L/min to 15 L/min) (HIGH); 0 to 10 L/min
Minute Volume Alarm
(LOW)
Continuous Pressure Alarm (PEEP + 1.5 kPa) over 16s
Suffocation Warning 5s to 60s no spontaneous ventilation
Maximum Limited Pressure 12.5 kPa
Fan Error Show On Screen
Oxygen Deficit Show On Screen
Gas Source O?, N?O
Pressure 280 kPa to 600 kPa
Power Supply 100V to 240 V
Dimensions (L × W × H) 740 × 800 × 1460 mm
Net Weight 66 Kg
Gross Weight 98 Kg
156 Great Charles Street, Queensway, Birmingham, West Midlands B3 3HN, Uk
Email: [email protected] | Website: www.medicaldeals.com | Phone: +1 506 699 0674