Labexpo Biosafety Cabinet Class II B2 ensures 100% exhaust with no recirculation, making it ideal for handling volatile chemicals and radionuclides. Featuring a HEPA filter, a vertical glass door, and an energy-efficient design, it guarantees top-tier protection for personnel, products, and the environment.
Stainless Steel Class II Biosafety Cabinet
Biosafety Cabinet Class II B2 56-BSC200
Biosafety Cabinet Class II B2 56-BSC200
Biosafety Cabinet is a Class II type B2 safety level 2 enclosed cabinet unit where there is no recirculation inside the workspace; 100
% air is exhausted as they are hard connected to exhaust system. This is used to handle volatile toxic chemicals and radio
nucleotides where chemical vapor re-circulation into the work zone is not permitted. It meets the requirements for the protection of
product, personnel and the environment allowing for the safe handling of pathogens, contaminants or other potentially hazardous
materials.
Features:
100 % filtered air exhaust
Vertical glass door / 10° tilted glass door
Superior filtration : 0.3 micron HEPA filter
Vertical negative pressure
Air insulation to avoid cross pollution
Microprocessor controller LCD digital display
Energy efficient
Stainless steel working space construction
Specifications :
Air exhaust 100%
Glass door orientation Vertical
Working area 1340 x 620 x 590 mm
Overall dimension 1540 x 785 x 1950 mm
HEPA filter 0.3 micron
Air pressure Vertical negative
Cleaning level 100 level
Ascending wind speed ? 0.55 m/s
Descending wind speed ? 0.35 m/s
Noise ? 62 dB
Half ventilation peak value ? 5 µm
Personnel protection CFU ? 10
Product protection CFU ? 5
Cross contamination CFU ? 2
Fluorescent lamp 16 W x 2
UV lamp 15 W x 2
Power AC 220 V / 50 Hz
Air supply filter dimension 1335 x 600 x 38 mm
Weight 250 kg